Description
Up to 5 points push-button calibration with auto-buffer recognition. selectable pH buffer standards, resolution and temperature units. Automatically calculates and displays electrode slope that helps you decide whether to replace pH electrode. ORP Mode: Single point offset calibration allows user adjusts displayed value to a known standard. Absolute or Relative mV measurements provides accurate readings Ion Mode: Standard BNC connector can connects many Ion selective Electrode. 2 to 5 points push-button calibration from low to high concentrations. selectable multiple concentration units, including ppm, mg/L, mol/L. Direct readout ion concentration simplifies measurement process. mV measurement allows user to check performance of ion electrode. Conductivity/TDS/Salinity Mode: Up to 5-point push-button calibration allows user to uses customized calibration solutions. Adjustable TDS factors, cell constants, temperature coefficients and normalization temperature. selectable linear or pure water compensation modes. In salinity measurement modes, selectable practical salinity (PSU) or seawater measurement standard (ppt). Dissolved Oxygen Mode: 1 or 2 points auto-calibration (100% in saturated air or air-saturated water, 0% in zero oxygen solution). Manual salinity and barometric pressure compensation. selectable readout in ppm, mg/L or percentage saturation. PH900 multi-parameter meter is capable of measuring up to 9 water quality parameters, including pH, ORP, Ion, Conductivity TDS, Salinity, Resistivity, Dissolved Oxygen, etc.
Features