Description
The PCR method is extremely sensitive, requiring only a few DNA molecules in a single reaction for amplification across several orders of magnitude. Therefore, adequate measures to avoid contamination from any DNA present in the lab environment (bacteria,viruses, or human sources) are required. Because products from previous PCR amplifications are a common source of contamination, many molecular biology labs have implemented procedures that involve dividing the lab into separate areas. One lab area is dedicated to preparation and handling of pre-PCR reagents and the setup of the PCR reaction, and another area to post-PCR processing, such as gel electrophoresis or PCR product purification. For the setup of PCR reactions, many standard operating procedures involve using pipettes with filter tips and wearing fresh laboratory gloves, and in some cases a laminar flow cabinet with UV lamp as a work station.
Features
ADVANTAGE:
1. LED display.
2. UV sterilization system.
3. HEPA filter efficiency 99.999% at 0.3μm.
4. Interlock function: UV lamp only can be switched on when the front window is closed, ensuring operator safety.
5. UV timer (1-99 minutes): When the setting time expires, the UV lamp automatically switches off in preparation
for the next experiment.
Standard Accessory:
UV lamp 3pcs
Fluorescent lamp 1pc
Base stand 1pc
Waterproof socket 2pc
Shelf with IV bar 1pc